Remote Control Commands
R&S
®
NRX
192
User Manual 1178.5566.02 ─ 07
.
Usage:
Event
DISPlay:MESSage:TEXT[:DATA]
<string>
Defines the text for user-defined messages.
Parameters:
<string>
ASCII characters
String "\n" causes a line break. The max. number of lines
depends on the message type: 9
lines for messages and 7
lines
for queries.
The length of a line depends on the characters used. Too long
lines are cut off.
DISPlay:MESSage:TYPE
<type>
Sets the message type for the user-defined messages.
Parameters:
<type>
QUERy | MESSage
QUERy
The execution of remote control commands is blocked, until the
dialog containing the query is closed.
MESSage
Remote control command processing is immediately continued.
Close the dialog containing the message by pressing [Esc] or
using
OFF
.
*RST:
MESSage
DISPlay:MESSage[:STATe]
<state>
If enabled, displays a dialog containing a user-defined message.
.
Parameters:
<state>
DISPlay:OVERload[:STATe]
<state>
If enabled, hides the sensor overload message. But be aware that overload can dam-
age the power sensor, depending on the amount of power and the duration of the over-
load condition.
You can query the allowed maximum power using
, or look it up in the data sheet of the power sensor.
Replaces the following R&S NRP2 command:
SERVice:DISPlay:OVERload
Measurement Settings and Results
www.allice.de
Allice Messtechnik GmbH