Command reference
R&S
®
ZNA
1285
User Manual 1178.6462.02 ─ 20
CPUosm
PFNPort (PUOSM) Scalar Vector Mixer calibration
<Characterization>
Location of the characterization (cal kit file) to be used for the
automatic calibration (string parameter):
–
If an empty string (
' '
) is specified, the factory cal kit file
stored in the active calibration unit is used.
–
A VNA cal kit file name
*.calkit
without path refers to a
specific cal kit file stored in the internal memory of the active cal-
ibration unit.
–
A VNA cal kit file name without path but prefixed with "SD:"
refers to a specific cal kit file stored on the SD card inserted at
the active calibration unit.
–
A VNA cal kit file name
*.calkit
with path refers to a spe-
cific cal kit file stored in an arbitrary directory on the analyzer.
<TestPort1>
Test port number. For an n-port automatic calibration, n arbitrary
(not necessarily consecutive) port numbers must be specified.
For a one path two port calibration (OPTPort), the first port no.
denotes the source port (fully calibrated port).
<CalUnitPort1>
Port numbers of the cal unit. For an n-port automatic calibration,
n arbitrary (not necessarily consecutive) port numbers must be
specified. It is possible to combine arbitrary (not necessarily
matching) pairs of analyzer and cal unit ports.
<TestPort2>
Second test port number
*RST:
n/a
Example:
CORR:COLL:AUTO:PORT FNPort, '', 1, 2, 2, 4, 4,
1
Perform an automatic full 3-port calibration at test ports 1, 2, and
4 using the calibration unit's default calibration kit file and ports
2, 4, and 1 of the cal unit.
See also
[SENSe<Ch>:]CORRection:COLLect:AUTO:
Usage:
Setting only
Manual operation:
See
[SENSe<Ch>:]CORRection:COLLect:AUTO:POWer
<TestPort>
Initiates a receiver power calibration at physical port
<TestPort>
. This is the second
step of an automatic SMARTerCal.
Note:
The power calibration must be performed "after" the system error correction; see
example below.
Suffix:
<Ch>
.
Channel number of the calibrated channel
SCPI command reference