SENSOR MAINTENANCE
RAIN GAUGE
Inspect and clean every two to three
months - removing leaves, debris, etc.
In extreme environmental conditions it
may be necessary to do this on a
MONTHLY basis.
•
Check that the tipping mechanism
moves freely and that it registers
0.01 inches per tip.
•
Check and Re-Level the Rain
Bucket, if necessary.
•
Calibrate the Rain Gage every 2
years.
Calibrating the Rain Bucket:
•
Secure a metal can that will hold at
least one quart of water.
•
Punch a
VERY SMALL HOLE
in
the bottom of the can.
•
Place the can in the top of the funnel,
of the Rain Bucket, and pour 16 fluid
ounces (1 pint) of water into the can.
•
If it takes
LESS
than 45 minutes for
this water to run out, the hole in the
can is
TOO LARGE.
•
One hundred (100) tips plus or minus
three tips should occur and register
from 0.97 inches to 1.03 inches.
(0.01 inches per each tip).
Adjusting screws are located on the
bottom of the Rain Bucket, adjacent to
the large center drain hole. Adjust both
screws the same number of turns or
partial turns. Rotation clockwise
increases the number of tips per 16
ounces of water, counter clockwise
rotation decreases the number of tips per
16 ounces of water.
One half turn
of
both screws causes a
2% to 3% change
.
•
Check and Re-Level the Rain Bucket
lid.
Factory Calibration is available through
your Rain Bird Distributor.
Suggestions:
Devices for dripping water are available
through a chemistry equipment supply
house.
In order to see the 100 tips, configure a
FAKE weather station within the
weather software, to use for this testing.
Drip the water and then do an upload
using the new test configuration to see
the number of tips (0.01 inches = 1 tip).
CAUTION !
Be sure to
disable
any
Automatic Uploads
, since the 100 tips
will contaminate the every day weather
data.
After the calibration is complete, reset
the datalogger by removing the power
and the battery connections in the
Enclosure.
Contact Campbell Scientific
for further instructions.
You will have
to reset the time of day and year in the
datalogger by up loading one day of data
(#6 from the Weather Main Menu). The
7 days of data that are stored in the
datalogger will be lost when you
disconnect the power, so be sure to have
done an upload before you begin the
calibration process.