PROGRAMMING
4 - 36
4.8 Change Sensor Configuration
In the programming mode, users may change several
sensor-related configurations of the
Q-RAE
monitor.
Table 4.7 shows the sub-menus for setting various
configurations of the sensors.
Table 4.7
Sensor Configuration Sub-Menu
Change LEL Gas Selection?
Enable / Disable Sensors?
Change Dilution Ratio?
Before these sub-menus are described, the term
“Correction Factor” needs to be explained:
Correction Factor
The LEL sensors used in Q-RAE
respond to a broad range
of gases that have different sensitivities. The correction
factor for a specific gas is defined as:
Sensitivity to a calibration gas
Correction Factor =
________________________
Sensitivity to a specific gas
Correction factors are often used to obtain calculated
concentrations of a specific gas while using a different gas
during calibration (typically methane or pentane).