Procedure:
•
After starting a test control following information data are displayed
in the display workspace:
[1] Product test
- Product code and name
Qn
- Product nominal mass
Load empty: 1/10
- Command concerning putting subsequent products
that are numbered out of the whole charge
1/30
- Number of performed full measurements out of the
sample quantity
TARE VALUES
COLLECTING
- Message for command „Load empty”
T
- Number of errors T (Lower than Qn –T) in the sample
2T
- Number of errors 2T (Lower than Qn –2T) in the sample
av.
- Average mass of controlled goods
OK
- Test results: OK. - positive, WRONG - negative
Test in progress
- Status
F5-End
- Description of button terminating control
NET
- Net mass of tested products
•
According to the command on the screen put on the weight pan
subsequent empty or full packages and after stabilizing press
,
•
After completing a test control different data are displayed in the
workspace (see ch. 18.7).
18.9. Full – empty control
For controls„full - empty” in computer software KTP NET in the product
settings users can set a measurement „charge”. The weighing program
during testing requires putting a charge of full packages and then a
charge of the same packages emptied put in the same sequence.
Procedure:
•
After starting a test control following information data are displayed
in the display workspace:
46