16.5. Software version
Procedure:
•
Enter
submenu
<
Others> according to ch. 16 of this manual,
choose <
Software Version> to see the program version
number.
16.6. Factory number
Procedure:
•
Enter
submenu
<
Others> according to ch. 16 of this manual,
choose <
Factory number> to see the program version number.
17. CUSTOMER CALIBRATION
An option only for non-verified scale
Scales require to recalculate internal divisions to more suitable ones (e.g. g,
kg etc.). In order to do this they require a calibration factor. It is adjusted
during the calibration procedure using a mass standard. Calibration should
be made when weighing a standard mass shows a different mass value.
To enter <
Customer Calibration>,
press
and then
:
<
Customer Calibration>.
17.1. Calibration procedure
•
Enter
submenu
<
Customer Calibration> according to ch. 17
and select: “
Platform 1 /
Calibration”,
•
After entering the parameter the following message box appears:
35