19 

 

8.4.3. Correction of Start Mass Expressed in Converter's Divisions 

 

  Enter factory setup submenu: 

<P.0.FAct / 0.3.CAL>

  Enter 

<0.3.4.Stu>

 submenu, value of start mass expressed in converter's 

divisions is displayed. 

  Correct  the  value,  to  do  it  use  operation  panel,  next  press 

  key  to 

confirm changes.  

  Exit to the home screen, to do it press 

 key repeatedly. 

 
 

8.5. Linearity Correction 

 

Prior  to  linearity  correction  it  is  necessary  to  determine  real  scale 
characteristics.  Correction  mechanism  allows  to  enter  corrections  in  20  points 
maximum.  

 
 

8.5.1. Entering Linearity Correction Points 

 

  Enter  factory  setup  submenu: 

<P.0.FAct  /  0.4.Line  /  0.4.1.dSG>

,  text 

<Cont?>

 is displayed. 

  Press 

 key to confirm. Text 

<Pnt1>

 is displayed (linearity connection 

point no. 1). 

  Press 

  key,  window  for  entering  mass  value  for  linearity  correction 

point no. 1 is displayed. 

  Enter  the  given  value,  next  press 

  key  to  confirm,  text 

<Pnt2>

  

is displayed (linearity connection point no. 2). 

  Press 

  key,  the  program  automatically  suggests  value  for  the  next 

linearity correction point.  

  Press 

  key  to  confirm  the  suggested  value,  or  enter  a  new  value  

(do it using the operation panel). 

  Keep  repeating  this  procedure  until  you  enter  value  close  to  maximum 

capacity value. 

 

 

While attempting to enter correction point value greater than 
maximum capacity value, <Err Hi> error is signalled. 

 

Содержание PUE H315

Страница 1: ......

Страница 2: ...2 SEPTEMBER 2021...

Страница 3: ...e immediately unplug the device from the mains Scales to be decommissioned must be decommissioned in accordance with valid legal regulations Do not let battery discharge in case of prolonged storage o...

Страница 4: ...ETUP 14 8 1 Access 14 8 2 Factory Parameters List 15 8 3 Device Defining 17 8 4 Factory Adjustment 18 8 4 1 External Adjustment Process 18 8 4 2 Start Mass Determination 18 8 4 3 Correction of Start M...

Страница 5: ...ized with assistance of manufacturer and user representatives C RADWAG does not bear any responsibility for damage or losses resulting from unauthorized or inadequate performing of production or servi...

Страница 6: ...of dishwashing detergent 3 2 Cleaning ABS Components To clean dry surfaces and avoid smudging use clean non colouring cloths made of cellulose or cotton You can use a solution of water and detergent s...

Страница 7: ...SB connector 5 RS232 1 connector 6 Universal socket or gland RS232 2 or RS485 or Ethernet or IN OUT 4 3 Pins Arrangement RS232 1 RS232 2 Pin1 NC Pin2 RxD Pin3 TxD Pin4 NC Pin5 GND Pin6 5VDC Pin7 NC Pi...

Страница 8: ...on key press to change the working mode Press to send the weighing result to a printer or computer Press to zero the scale Press to tare the scale Upon pressing keys combination functions of given key...

Страница 9: ...s with shield Max number of platforms 1 Multirange Yes RS232 1 M12 8P connector USB M12 4P connector Optional Equipment RS232 2 M12 8P connector RS485 M16 cable gland Ethernet M12 4P connector 4 20mA...

Страница 10: ...Pictogram displayed continuously Too low battery charge the scale is about to shut down Charge the battery immediately Blinking pictogram blink frequency ca 1 s Battery charge in progress The device i...

Страница 11: ...current parameter value Press to enter given submenu Press to modify given parameter Press to confirm modification Press to exit function remains unmodified Press to move one menu level up 6 1 Return...

Страница 12: ...oad cell socket Load cell signals NOTES REF SENSE JP3 not soldered REF SENSE JP4 not soldered IN OUTPUT IN OUTPUT 5V INPUT AGND INPUT 7 2 Connecting 4 Wire Load Cell Connect 4 wire load cell to the ma...

Страница 13: ...Cable Shield Platforms connected to indicators in metal housing via a cable only Platforms electrically connected to indicators metal housings e g pillars racks Load cells with internal shield connect...

Страница 14: ...h the weighing device off to do it press key Press and hold S1 switch located on the electronics board while holding the S1 switch press key to run the weighing device Access to factory setup switch W...

Страница 15: ...c RTC synchronisation 0 1 A ntE YES no US market metrological requirements activation 0 2 nnG Metrology 0 2 1 A d Converter s divisions 0 2 2 Uni g kg lb Adjustment unit 0 2 3 du1 0 0001 0 0002 0 0005...

Страница 16: ...mass control YES range from 10 to 10 of start mass no off 50 range from 50 to 50 of start mass dEF range declared in 0 2 G parameter 0 2 G rnt 10 90 Start mass range in 0 2 H Ldn no YES Digit marker...

Страница 17: ...parameters serial number weighing device type weighing range Procedure Enter factory setup P0 FAct Go to 0 1 Glob 0 1 1 duu submenu text Cont is displayed Press key text nr FAc is displayed next you s...

Страница 18: ...ighing pan with the required adjustment weight Press key adjustment starts After adjustment completion 0 3 1 CLE submenu is displayed automatically Exit to the home screen to do it press key repeatedl...

Страница 19: ...Correction Points Enter factory setup submenu P 0 FAct 0 4 Line 0 4 1 dSG text Cont is displayed Press key to confirm Text Pnt1 is displayed linearity connection point no 1 Press key window for enter...

Страница 20: ...ext linearity correction point to do it press key Correcting the linearity correction point no 2 is analogous In order to finish correcting linearity correction points press key 0 4 3 Cor submenu is d...

Страница 21: ...o be mounted inside the indicator It is equipped with optoisolated inputs and semiconductor outputs and enables free configuration of both the inputs and the outputs using indicator menu The indicator...

Страница 22: ...odule 9 1 2 I O Schematic Diagrams 4 inputs 4 outputs 9 1 3 Input Output Signals Overview INPUTS OUTPUTS Wire number Signal Wire number Signal 1 IN1 6 OUT1 2 IN2 7 OUT2 3 IN3 8 OUT3 4 IN4 9 OUT4 5 COM...

Страница 23: ...an additional cable gland through which 3 metre cable 2 x 0 25 mm2 is fed which cable is terminated with isolated wires 4 20mA 636R module 9 2 1 Technical Specifications Output resolution 16bit Linea...

Страница 24: ...nds the PUE H315 indicator functionality by RS485 interface The module is intended to be mounted inside the indicator The indicator s housing cover features an additional cable gland through which 3m...

Страница 25: ...over features an additional M12 4P D codded connector Ethernet standard RS485 635R module 9 4 1 Ethernet Pins Ethernet Pin1 RX Pin2 TX Pin3 RX Pin4 TX 9 5 Additional Modules Arrangement 635R board inc...

Страница 26: ...26 Exemplary assignment of modules on 634R board Additional modules arrangement 10 DIAGRAMS OF CONNECTION CABLES Indicator computer cable Indicator printer cable EPSON...

Страница 27: ...27 Indicator USB adapter cable...

Страница 28: ...28...

Отзывы: