18
While defining balance/scale type, additional parameters are
set automatically. These are among many battery type,
internal adjustment accessibility, additional modules and
communication interfaces accessibility.
8.4. Factory Adjustment
8.4.1. External Adjustment Process
Enter factory setup submenu:
<P.0.FAct / 0.3.CAL>
.
Enter
<0.3.1.CLE>
function, text
<UnLoAd>
is displayed.
Unload the weighing pan.
Press
key, adjustment zero point gets determined.
With adjustment zero point determined, first text
<LoAd>
, next mass of an
adjustment weight that is to be loaded onto the weighing pan is displayed.
Load the weighing pan with the required adjustment weight.
Press
key, adjustment starts.
After adjustment completion,
<0.3.1.CLE>
submenu is displayed
automatically.
Exit to the home screen, to do it press
key repeatedly.
8.4.2. Start Mass Determination
Enter factory setup submenu:
<P.0.FAct / 0.3.CAL>
.
Enter
<0.3.2.Std>
function, text
<UnLoAd>
is displayed.
Unload the weighing pan.
Press
key, start mass determination starts.
Upon completed determination process,
<0.3.2.Std>
submenu is
displayed.
Exit to the home screen, to do it press
key repeatedly.
If
start
mass
determination
or
adjustment
factor
determination takes more than 360 seconds then <Err8>
error is displayed and short sound signal is heard. In such
a case the scale requires readjustment/recalibration,
perform it providing as stable ambient conditions as
possible!