![RADWAG PUE C32 Скачать руководство пользователя страница 25](http://html1.mh-extra.com/html/radwag/pue-c32/pue-c32_user-manual_3385121025.webp)
25
9.4. Return to Weighing
Introduced modifications are automatically recorded upon return to the home
screen. To return to the home screen:
press
key repeatedly, keep pressing the key until you see the
home screen,
press
key, the home screen is displayed immediately.
9.5. Factory Adjustment
9.5.1. External Adjustment
Enter
<Factory/Adjustment>
submenu.
Select
<External adjustment>
parameter. Message
<Remove weight>
is displayed.
Unload the weighing pan and press
key to confirm.
Message
<Adjustment. Please wait...>
is displayed.
Upon completed start mass determination procedure, message
<Put weight xxx>
is displayed (where: xxx
– adjustment weight mass).
Load the weighing pan with a required adjustment weight and press
key to confirm. Message
<Adjustment. Please wait...>
is displayed.
Upon adjustment completion, message
<Remove weight>
is displayed.
Unload the weighing pan, the weighing device displays the
<Adjustment>
submenu.
9.5.2. Start Mass Determination
Enter
<Factory/Adjustment>
submenu.
Select
<Start mass determination>
parameter. Message
<Remove
weight>
is displayed.
Unload the weighing pan and press
key to confirm.
Message
<Start mass determination. Please wait…>
is displayed.
Upon completed determination process, the weighing device displays
the
<Adjustment>
submenu.
9.6. Linearity Correction
9.6.1. Linearity Determination
Declaring mass values for subsequent linearity steps and determining