21
On RADWAG board
Load cell signal
Notes
E
SHIELD
See section 9.3
REF+
-
JP3 soldered
REF-
-
JP4 soldered
IN+
IN-
OUTPUT-
+5V
INPUT+
AGND
INPUT-
8.3. Load Cell Shield Connection
Load cell with galvanically
connected shield of the
signal cable.
Load cell w/o galvanically
connected shield
of the
signal cable.
Scale with an indicator in an ABS
housing connected with a platform
via a load cell signal cable.
DO NOT connect
E
Compact scale design (scale with
an ABS indicator on a post, etc.)
DO NOT connect
E
E
– Soldering spot on the main board and on the additional A/D converter boards.
9. FACTORY SETUP
To access and modify factory settings and parameters that are made available
for a user, run Factory Setup mode. Running Factory Setup mode enables the
technician to define the scale.
9.1. Factory Setup Access
Switch the weighing device on, to do it press
key.
Upon completed start-up, weighing result window is displayed
automatically.
Using pin, screwdriver, hex key, etc., press an internal
S1
switch, to do
it you must find an opening of Phi 3.2 mm in the bottom of the indicator
base, insert the pin through it and press the switch.