23
Service
-
-
Service settings menu.
Battery diagnostics
-
-
Preview of operation parameters and
battery charge-up.
RTC adjustment
-
-
RTC synchronisation.
Operation time
-
-
Total operation time, time format:
"D: xx H: yy" (D - day, H - hour).
CPU Id
-
-
Processor
number,
necessary
to
generate code unlocking Wi-Fi module.
Restore default
operator settings
-
-
Restoring default settings (factory and
user parameters).
9.3. Factory Parameters
NAME
DEFAULT
VALUES
DESCRIPTION
Metrology
-
-
Metrology related settings
Mass divisions
-
-
Display of converter divisions.
Adjustment unit
kg
g, kg, lb
Adjustment unit.
GCOR coefficient
1
0.9
1.1
Gravitational correction factor.
Reading unit - Range 1
0.001
0.0000001
50
Reading units for Range 1.
Verification unit
- Range 1
None
None, 0.001, 0.002,
0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Verification units for Range 1, 'none'
value - unverified version.
Reading unit - Range 2
0.001
0,0000001
50
Reading units for Range 2.
Verification unit
- Range 2
None
None, 0.001, 0.002,
0.005, 0.01, 0.02,
0.05, 0.1, 0.2, 0.5,
1, 2, 5, 10, 20, 50
Verification units for Range 2, 'none'
value - unverified version.
Digit marker
None
None, 1, 2, 3
Setting quantity of digits (counting
from the last digit of the displayed
result) over which a marker is to be
displayed and printed out in case of
unverified scales.
Max range
3.009
-
Maximum ca overload.
Range 2 threshold
0.000
-
II range switching point. For single
range scales set 0.
External adjustment
weight
3.000
-
Mass of an external adjustment
weight.
Autozero range
Predefined
Predefined, 0.1, 0.2,
0.25, 0.5, 0.6, 0.7,
0.8, 0.9, 1, 2, 2.5, 3,
4, 5, 6, 7, 8, 9, 10.
Autozero range [d]: Predefined -
value
taken
from
program-
implemented tables; 0.1
10 - value
entered directly by a user.