Set respective parameter values and return to weighing, saving changes.
11.3.2. Defining Transmission Parameters for HRP Weighing Module
In order to define transmission parameters of HRP weighing
module, press an internal SW1 switch on the main board
of the module. Follow HRP service manual.
Make sure that communication between the PUE 7.1 and the HRP is
established, enter factory parameters submenu of the PUE 7.1 indicator.
Go to a modified submenu of the weighing platform.
Enter
„Miscellaneous / MW-MH: Communication”
submenu, list of the
following parameters is displayed:
NAME
DEFAULT
VALUES
DESCRIPTION
RS232 (1)
57600
1200
115200
Baud rate of RS232 (1) port of HRP
weighing module.
RS485
57600
1200
115200
Baud rate of RS485 port of HRP
weighing module.
Address
1
-
Setting address of HRP module.
IP Address
192.168.0.2
-
Setting IP address of HRP module.
Subnet mask
255.255.255.0
-
Setting subnet mask of HRP module.
Default gateway
192.168.0.1
-
Setting default gateway of HRP module.
Port
4001
-
Setting local port of HRP module.
Timeout
60
0
300s
Setting time interval after passage
of
which
an
inactive
Ethernet
connection gets broken.
Set respective parameter values and return to weighing, saving changes.
In order to save changes in HRP module memory it is
necessary to restart PUE 7.1 indicator.
11.3.3. Defining Factory Settings for HRP Module
In order to determine factory settings of HRP weighing
module press an internal SW1 button on the main board of
the module. Follow HRP service manual.
Make sure that communication between the PUE 7.1 and the HRP is
established, enter factory parameters submenu of the PUE 7.1 indicator.