10.3. Factory Parameters
NAME
DEFAULT
VALUES
DESCRIPTION
Metrology
-
-
Metrology-related settings.
Reading unit: range 1
0.001
0.001
50
Rounding of the last digit and position of
a decimal point for weighing range I.
Reading unit: range 2
0.001
0.001
50
Rounding of the last digit and position of
a decimal point for weighing range II.
Reading unit: range 3
0.001
0.001
50
Rounding of the last digit and position of
a decimal point for weighing range III.
Range
6.009
-
Maximum ca overload.
Range 2 *
0.000
-
Switching point of range II.
Range 3 **
0.000
-
Switching point of range III.
Adjustment weight
mass
2
-
Adjustment weight mass.
Adjustment unit
kg
g, kg, lb
Adjustment unit.
g-cor.
1
-
Gravitational correction value.
Autozero range
0.25
0.1
5
Autozero range (in divisions).
Stability range
1
0.1
5
Stability range in [d] units.
Stability time
0.8
0
20
Stability time in [s].
Start mass control
1
0, 1, 2
Start mass control: 0
– disabled, 1 – range:
from -5% to +15% of start mass, 2
– range:
from -20% to +20% of start mass,
Digit marker
No
No,
Yes
Enabling/disabling digit marker.
Adjustment
-
-
Adjustment process (read section 10.5).
Start mass
determination
-
-
Determining platform start mass.
Adjustment
-
-
Platform calibration.
Adjustment with pan
loaded
-
-
Loaded platform calibration.
Start mass
-
-
Start mass value (in divisions).
Adjustment factor
100
-
Adjustment factor.
Adjustment weight
mass
Adjustment weight mass.