4. UNPACKING, ASSEMBLING AND STARTUP
4.1. Assembling and place of use
•
Balance should be stored and used in locations free of vibrations and
shakes, free of air movement and dust.
•
Ambient air temperature should not exceed the range of: +10
o
C
÷
+40
o
C
•
Ambient relative humidity should not exceed 80%.
•
During balance operation, ambient temperature in the weighing room should
not change more than 3°C within one hour,
•
The balance should be located on a stable wall console desk or a stable
working table which is not affected by vibrations and distant from heat
sources.
•
Take special safety measures when weighing magnetic objects, as part of
the balance is a strong magnet. Should such loads be weighed, use under
hook weighing option, which removes the weighed load from area influenced
by the balance’s magnet. For assembling the hook for under hook weighing
option see the bottom section of balance’s housing.
In order to avoid influence of static electricity on the measurement process,
ground the balance’s housing. the grounding bolt is located at the back sof
balance’s housing.
4.2. Unpacking
Carefully remove the balance from its packaging, remove the plastic and foil
transport protective elements and gently place the balance in its intended place
of use. Assemble the weighing pan, and other elements according to below
schema:
Содержание PS 1000/C/2
Страница 2: ...2 JUNE 2012...
Страница 6: ...6...
Страница 30: ...30 The higher filtering setting the longer weighing time...
Страница 37: ...37 100 maximum brightness of display backlight 20 minimum brightness of display backlight no backlight off...
Страница 60: ...60 During setting limit values the following relations occur...
Страница 70: ...70...
Страница 85: ...85 should be omitted as they are added only for the purpose of proper legibility...
Страница 90: ...90...