75
15.3. Setting operation mode for RS 232 interface
no
–
manual operation (enabled on pressing
key)
YES
– automatic operation (enabled on stabilization of weighing
result).
Return to weighing
(see point – 8.2. – return to weighing).
Automatic operation takes place according to the following scheme:
press
Esc/TARE
key to zero the balance (display will indicate
pictograms of stable measurement
and precise zero
)
place the load, balance will send first stable measurement through the
RS232 interface
remove the weighed load from the weighing pan,
next measurement can be conducted if a weighing result /- 50
reading units from zero (it is not necessary to reach precise zero to
start another measurement).