- 81 -
•
Assign <Checkweighing thresholds> option to one of the quick access
keys displayed in the bottom bar
•
Return to drying mode
•
Press <
Checkweighing thresholds> soft key and insert the value of
Low Threshold and High Threshold. Return to drying mode.
•
After displaying <PREPARE SAMPLE> command, the display, below the
measurement result, indicates a bar graph, which using colours shows
current mass on the drying pan.
•
Yellow colour – current mass below set Low Threshold value
•
Green colour – current mass within set thresholds values
•
Red colour – current mass above set value of High Threshold
2.
Defining the value of target mass is carried out using a quick access key
<
Target mass >
Procedure
•
Press grey coloured workspace, and then <Keys> option
•
Assign <Target value> option to one of the quick access keys displayed
in the bottom bar
•
Return to drying mode
•
Press <
Target value> soft key and insert mass to be obtained in
measuring process,
•
Then, /- percent tolerance of the target mass, (accessible range
0 – 100%)
•
The display, below the measurement result, indicates a bar graph
which contains:
o
Current mass (visualized with colours)
o
Target Mass value (visualized with black marker)
Содержание MA 3Y