31.
MOISTURE ANALYZER MAINTENANCE ACTIVITIES
This section of the user manual describes procedures of maintaining a moisture
analyzer MA 3Y series in good condition, and means of replacing faulty
components (fuses).
31.1.
Cleaning moisture analyzer components
In order to ensure required measuring accuracy, the moisture analyzer has to be
used and stored in clean conditions.
While cleaning the moisture analyzer, precautions provided in this section of the
user manual must be obeyed.
REMEMBER,
before initiating any maintenance or cleaning
activities of moisture analyzer’s inner parts, switch off the moisture
analyzer by pressing ON/OFF key on terminal’s overlay, and make
sure the power cord of the moisture analyzer is unplugged from
mains!
Open moisture analyzer’s lid and take
out all components of the drying
chamber:
disposable pan (5),
cross-shaped holder (4),
pan handle (3),
drying pan shield (2),
insert of the drying chamber base (1).
Clean the moisture analyzer using soft
fabrics (diaper, chamois, etc.) and
non-aggressive cleaning agents.
Do not use any abrasive agents or
solvents to clean the moisture
analyzer, as it may cause damage to
the device and its components.
After cleaning, thoroughly dry each of moisture analyzer’s components.
Do not allow any contamination or liquid to fall into the moisture analyzer.
Assembly cleaned and dried components in accordance with above scheme.