- 139 -
Send stable measurement result in basic measuring unit
Format:
S CR LF
Accessible responses:
S_A CR LF
S_E CR LF
- command understood and in progress
- time limit exceeded while waiting for stable
measurement result
S_I CR LF
- command understood but not accessible at this
moment
S_A CR LF
MASS FRAME
- command understood and in progress
- response is mass value in basic measuring unit
Mass frame format which is moisture analyzer’s response:
1 2-3
4
5
6
7-15
16 17 18 19 20 21
S space stability
marker space character mass space
unit
CR LF
Example:
S CR LF
– command sent from a computer
S _ A CR LF
- command understood and in progress
S _ _ _ _ - _ _ _ _ _ _ 8 . 5 _ g _ _ CR LF
- command carried out, response
is mass value in basic measuring unit.
Immediately send measurement result in basic measuring unit
Format:
SI CR LF
Accessible responses:
SI_I CR LF
- command understood but not accessible at this
moment
MASS FRAME
- response is immediate with mass value in basic
weighing unit
Mass frame format which is moisture analyzer’s response:
1 2
3
4
5
6
7-15
16 17 18 19 20 21
S I space stability
marker space character mass space
unit
CR LF
Example:
S I CR LF
– command sent from a computer
S I _ ? _ _ _ _ _ _ _ 1 8 . 5 _ g _ CR LF
- command carried out, immediate
response of mass value in basic weighing unit