Load againg standard weight “A” and the device enters the
stabilization process showing “-----“,
•
After about 2 minuts a weighing result appears on the display
additionally accompanied by a beep sound lasting for about
2 sekunds,
•
After the stabilization process the comparator performs a measurement
in mode „Automatic first stable”. (pictogram
appears),
•
Some descriptions on the display change. The standard to be weighted
changes to <A1> and the number of measurement cycles to <P 2/1 0>.
•
Repeat measurements as declared in parameter <
Number
of measurements >.
•
After the last measurement in a series is confirmed, the message
window is displayed:
•
After
pressing
a final report is printed out comprising
subsequent measurements and final calculations.
An example report:
84