Scale parameters - Weighing
Icon Description
Value
Platform
1
-
Median Filter
None
Filter Slow
Autozero No
LO threshold
0.1
Working modes - Comparator – Komparator
Icon Description
Value
Comparator
-
Save Mode
Automatic first stable
Checkweighing No
Tare mode
Single
Statistics Global
Method ABBA
Number of measurements
6
Ask for numbers of standards
No
3. Start the comparison process and follow directions appearing on the
display of indicator.
Notice:
Before you put the first weight on the pan you need to zero the device
manually by pressing button
on the indicator keypad.
4. Put on the pan the first standard weight “A”. On the display you will see
“-----“ signalling the process of stabilization of the measurement system.
81
Содержание KB series
Страница 2: ...2 MAY 2012...