background image

18

HR-501 User Manual (Ver.2.3)

Name of model

Purpose of use

Case type

Product size

Product weight

Free warranty

- HR-B1
- Storage and protection of AED from foreign 

-

 substances, storage of pads and manual

- Plastic, artificial leather

- 60(W) x 175(H) x 25(T)

- 370g

 
- 1 year (in case of damage caused by user’s 

-

 mistake, repair/exchange will be charged)

Name of model

Purpose of use

Pads type

Pads quality

Product size

Product weight

Product life

Free warranty

- P-303

- Electric shock and delivery of impedance signal

- Adhesive pads for both adult/kid

- Polyolefin form, adhesive elements: 180-129A

- 145(H) x 98(W) x 1(T)

-120g

 
- In case of storage of sealed package, 2 years from

date of manufacture; use immediately on opening 

packing (single-use)

- 2 years (in case of damage caused by user’s mistake,

  repair/exchange will be charged)., In case of more 

  than single-use, exchange with new one is essential

4.1.3 Accessories

HR-501,501T portable case

P-303 (HR-501 electric pads)

HR-B1 (HT-501, 501T)

HR-B1 (HT-501, 501T)

Содержание HR-501

Страница 1: ...er manual This user manual is keep users safety and to protect the damage on your property Do not lose user manual Please keep a user manual to be able to see at any time Copyright RADIAN Corp All Rig...

Страница 2: ...As a tracking medical device HR 501 requires registration of installation place and administrator Online registration is available in RADIAN s website Administrator of HR 501 must be well informed of...

Страница 3: ...5 3 Product maintenance and upgrades 6 Troubleshooting 6 1 Self diagnosis 6 2 Condition verification 6 3 Troubleshooting 7 Technical Data 7 1 BTE Biphasic Truncated Exponential wave form 7 2 Audio gui...

Страница 4: ...ds that may cause serious personal injury or death Reference additional reference data Disposable Single patient use only Do not recharge battery Additional information is provided in the AED The AED...

Страница 5: ...tton Non Sterile Use pads by this date Defibrillator protected by Type BF Upper and lower operating temperature limit Not water proof Date of manufacture year and month Waste Electronic Electrical Equ...

Страница 6: ...fore using the product it is very necessary to check the patient s consciousness and respiratory condition do not use if the patient is conscious and reacts Warning The AED s Basic Setting is Adult Mo...

Страница 7: ...Do not use under water When the patient is exposed to rain or snow use after the patient is dry in a dry area a When the patient is in contact with metals do not touch the metals Warning Do Not Use A...

Страница 8: ...t Damage the Battery BT 303 Recharge or heat damage of battery BT 303 may cause an explosion or fire hazard When the battery runs out exchange with RADIAN Inc or at a sales store Warning Do Not Disass...

Страница 9: ...fter removing any foreign substance between patient s skin and pads When pad is not attached to skin properly check if surface of pads is dry If surface of pads is dry then exchange with new pads Duri...

Страница 10: ...in case of long term storage it is encouraged to keep from 0 40 In case of electric pads there may be problem in product performance if it conducted long term use and storage out of range of 0 40 Warn...

Страница 11: ...surface of the AED Implement periodic management of parts life Warning The AED Requires Periodic Maintenance Control Maintenance control of the AED and authorization and responsibility of usage must b...

Страница 12: ...t defect of parts including battery and electric pads provided along with HR 501 The warranty is not applicable to devices transferred or traded Despite occurrence of product defect within warranty pe...

Страница 13: ...o call customer service at 82 1670 1980 for shipping method Shipping will be completed within a maximum of 30 days from the day the product arrives at RADIAN Inc which is automatically added up to the...

Страница 14: ...tion menu Heart Guardian HR 501 as a portable device operated by battery is an easy to use Semi automated External Defibrillator The AED identifies whether electric shock is needed by automatically an...

Страница 15: ...User Manual READY USE MANAGEMENT OTHER A E H I B C D J G 4 1 1 Product Overview Power button button Status indication light Handle Cover Shock button Pads connector link Procedure Indication Light Adu...

Страница 16: ...Basic setting is Adult mode Implements electric shock A link to connect pads connector to the product Announces the stage of current process in product use Send or receive data from a PC with Bluetoot...

Страница 17: ...tic diagnosis battery power circuit power button operation check NEWFM ECG Analysis Korean Chinese and English dubbing record Adult Kid compatible switch basic HI Cap charge 150J Adult 50J Kid 40 min...

Страница 18: ...lity Product size Product weight Product life Free warranty P 303 Electric shock and delivery of impedance signal Adhesive pads for both adult kid Polyolefin form adhesive elements 180 129A 145 H x 98...

Страница 19: ...eight Number of shocks Battery life Free warranty Warranty in charge HR 501 battery BT 303 Stable power supply Adhesive pads for both adult kid Plastic 60 H x 175 W x 25 T 285g More than 200 times Sta...

Страница 20: ...nc conducts education of CPR and AED in accordance with customer request Visit RADIAN s website www radian co kr and refer to the education menu 4 2 2 Subject of Product Use Heart Guardian HR 501 is a...

Страница 21: ...r proper use Open the packing box and confirm product components If the components are not prepared inquiry at the store Download of manual and quick guide is available at www radian co kr From the mo...

Страница 22: ...se 3 Insert battery pack into battery hole as shown in the picture below left Right Please check right direction Check user manual before operate Defibrillator Before setup the battery Check manufactu...

Страница 23: ...t is operating normally The AED turns off and shifts to standby status after conclusion of the audio message Self diagnosis ends In case of problem in equipment it informs about the problems by audio...

Страница 24: ...ectromagnetic intervention or turn off other electronic devices The devices that may cause electromagnetic intervention are cell phones radiophones and X ray The battery included in the product is not...

Страница 25: ...power by pushing power button Audio guidance Select adult kid switch with caution Heart Guardian HR 501 may be used for both adults and kids by button set up At the time of product delivery basic mode...

Страница 26: ...ult kid switch is placed on kid mode If adult kid switch is located in adult mode move adult kid switch to the kid mode as shown in the picture In case of unavailability of accurate weight and age of...

Страница 27: ...is very important for patients of sudden cardiac arrest Time must not be wasted to take off the clothes Remove patient s top as quickly as possible even by cutting or ripping it Make skin of patient s...

Страница 28: ...in the below picture Adult Children Infant CAUTION Despite switched placement of pads AED works properly Despite attachment in reverse make use of the product according to the audio instruction Conduc...

Страница 29: ...5 When electric shock is needed in case of need of defibrillation 6 Adult Children Infant Caution During ECG analysis stay away from patient Once ECG analysis is completed the product operates as sho...

Страница 30: ...guidance Conduct chest compressions 30 times Beeper rings 30 time Audio guidance Conduct artificial respiration 2 times Audio guidance Pause CPR Audio guidance Analyzing ECG Conduct CPR according to t...

Страница 31: ...n case of unnecessary defibrillation If electric shock is unnecessary the product operates as shown below Below is one cycle While the product conducts the CPR guide it does not analyze patient s ECG...

Страница 32: ...ing product use information In case of information delivery via IrDA attach the IrDA adaptor of the PC to the IrDA transmission port At the moment of pushing shock button push power button Audio guida...

Страница 33: ...irst of all have the product set maintenance mode to operate and complete Bluetooth search as well as transmission connection among PC and the product Thereafter install and run Heart Guardian GUI PC...

Страница 34: ...check periodically product operation and expiration date for battery and pads After it is used inspect the following issues Since the product is used in emergency all users must keep it ready to use...

Страница 35: ...35 HR 501 AED User Manual READY USE MANAGEMENT OTHER 2 Replace single use electric pads 3 Confirm battery manufacture date as well as remaining energy...

Страница 36: ...dity range 5 95 no saturated place 3 Maintenance precautions avoid product storage in the following places Avoid place that is vulnerable to temperature Avoid place that is exposed to direct light Avo...

Страница 37: ...ing dual holding switch of battery pack at the back of the product Push in the new battery pack facing the label below until clack sound emits Caution Do not use battery not provided by RADIAN RADIAN...

Страница 38: ...her maintenance parts are purchased users may confirm periodically if they are in order Users are recommended to check expiration date of pads periodically Pads with passed expiration date must not be...

Страница 39: ...ery second for warning with error audio guidance It implements every time battery is replaced The product conducts test at the time of initial battery installation after purchase or battery replacemen...

Страница 40: ...e the light blinks Third procedure the light button Red blink Shock button Orange blink prompts the state of proper use prompts the state of normal operation prompts when error is found during operati...

Страница 41: ...hock replacewithnewbattery Replace with new battery Push button according to audio guidance System error Error code is 0000 Contact vendor The product is turned off after 10 minutes or 10 times of ele...

Страница 42: ...e I Duration ms Phase II Duration ms Energy J 2 86 4 0 5 04 6 17 7 31 8 45 9 58 2 86 4 0 5 04 6 17 7 31 8 45 9 58 150 150 151 150 151 150 150 25 50 75 100 125 150 175 2 86 4 0 5 04 6 17 7 31 8 45 9 58...

Страница 43: ...hed and terminated Equipment abnormal state guidance Push blinking red i button Low battery Replace new battery immediately Replace battery soon System error Error code is 0001 low battery System erro...

Страница 44: ...n how to conduct CPR Conduct 30 times of chest compression Conduct 2 times of artificial respiration After every set of 30 chest compressions and two artificial respirations pause the CPR 7 2 5 Electr...

Страница 45: ...45 HR 501 AED User Manual READY USE MANAGEMENT OTHER...

Страница 46: ...2 2 1670 1980 CUSTOMER S E R V I C E RM1610 SIGMA VALLEY GASAN DIGITAL 2 RO 53 GEUMCHEON GU SEOUL 153 802 SOUTH KOREA T E L 82 2 6443 6600 F A X 82 2 6499 3076 W e b www radian co kr E_mail info radia...

Отзывы: