31
HR-501 AED User Manual
READ
Y
USE
MANA
GEMENT
O
THER
Audio guidance: ‘Electric shock is unnecessary.’
Audio guidance: ‘Identify respiration and consciousness.’
Audio guidance: ‘If there is neither respiration nor
consciousness, conduct CPR.’
•
•
•
Audio guidance: ‘Electric shock is not operated. Attach
pads to patient’s skin tightly.’
•
Audio guidance: ‘Shock button is not pushed.’
•
When electric shock is unnecessary (in case of unnecessary
defibrillation)
If electric shock is unnecessary, the product operates as shown
below.Below is one cycle.
While the product conducts the CPR guide, it does not analyze
patient’s ECG. Once CPR guidance is completed, the product
informs via audio message and reanalyzes ECG.
8)
Other conditions
A) In case the pads are attached to the skin incorrectly, the
product operates as shown below.
B) In spite of recharge completed, if the user does not push the
blinking shock button for more than 15 seconds, the product
does not deliver electric shock by pads but conducts
self-discharge.
9)