Overview 1-11
FM
The FM function modulates the 3155 sample clock frequency. You
can frequency modulate the output either with built-in waveforms, or
by downloading complex waveforms into FM modulation memory.
Using DDS technology, the modulation is wide band and extremely
linear. FM can be used in continuous, triggered and gated modes.
The 3155 can generate two types of frequency modulation: 1) Stan-
dard and 2) Arbitrary. In standard mode, the modulating waveform is
selected from a built-in library of 4 standard waveforms: sine, trian-
gle, square and ramped. In arbitrary mode, the modulating signal is
downloaded into FM modulation waveform memory. There are
100,000 points allocated specifically for arbitrary memory that are
used for arbitrary frequency modulation.
Output Type
The 3155 can output three types of waveforms: standard (Fixed), ar-
bitrary (User) and sequenced waveforms. A description of the vari-
ous waveform types that the instrument can generate is given below.
Standard (FIXED)
Waveforms
The 3155 must pre-load its memory before it can generate wave-
forms. On power up, the waveform memory has no specific data.
The sine waveform, being the default waveform at power on, is com-
puted and loaded into waveform memory as part of the reset proce-
dure.
Standard waveforms are using the same start address. Every time
that a new waveform is selected, there is some minimal time for the
processor to compute and download the data to the memory.
The 3155 can be programmed to output one of nine standard wave-
form shapes: sine, triangle, square, pulse/ramp, sine(x)/x pulse,
gaussian pulse, rising/decaying exponential pulse, noise and DC.
There are some parameters associated with each waveform, which
modify the shape of the waveform to better suit your needs. For ex-
ample, different start phases for the sine waveform can be pro-
grammed for each channel to create phase offsets between two or
more instruments.
Содержание 3155
Страница 4: ...This page was left intentionally blank...
Страница 14: ...User Manual 3155 1 2 Overview Figure 1 1 3155 Series...
Страница 16: ...User Manual 3155 1 4 Overview Figure 1 3 WaveCAD Wave Composer Figure 1 4 WaveCAD FM Wave Composer...
Страница 41: ...User Manual 3155 Installation 2 15 Figure 2 15 Device Manager...
Страница 44: ...User Manual 3155 2 18 Installation Figure 2 19 Customer Information Step...
Страница 110: ...User Manual 3155 3 58 ArbConnection This page was left intentionally blank...
Страница 150: ...User Manual 3155 Programming Reference 4 40...
Страница 184: ...User Manual 3155 Programming Reference 4 74 This page was left intentionally blank...
Страница 202: ...User Manual 3155 Specifications A 8 This page was left intentionally blank...