Publication No. 980855 Rev. A 1256 User Manual
EADS North America Test and Services Troubleshooting C-5
IEEE-488 (GPIB) Remote Interface
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
•
1256 is not set for remote
GPIB operation.
•
Use front-panel controls to set
the 1256 for GPIB operation.
See
Chapter 4, GPIB AND
RS-232 INTERFACES
, under
the heading
Selecting the
GPIB Interface
.
•
Commands sent to wrong
GPIB address.
•
Make sure that the commands
are sent to the GPIB address
currently assigned to the 1256
(to set the 1256 GPIB address,
see
Chapter 4, GPIB and RS-
232 Interfaces
, under the
heading
Setting the GPIB
Address
).
1256 does not respond to
GPIB commands.
•
The 1256 is set to the same
GPIB address as another
device in the system.
•
Make sure that the 1256 is set
to a unique GPIB address(to
set the 1256 GPIB address,
see
Chapter 4, GPIB and RS-
232 Interfaces
, under the
heading
Setting the GPIB
Address
).
Plug-In Modules
PROBLEM
POSSIBLE CAUSES
WHAT TO DO
Slot containing a plug-in is
indicated as an “empty slot”
when using front-panel
controls to scroll through
slots.
•
Plug-in not properly seated
in slot.
•
Loosen the retaining screws of
the plug-in. Pull the plug-in
part-way out, then gently push
it back in until it stops. Then
press it firmly to seat its
connector with the backplane
connector. When properly
installed, the panel of the plug-
in is flush with the rear panel of
the 1256.