5.
Turn the camera collar in a counter-clockwise direction, and loosen it
completely (see Figure B-1). The camera is now free to move up and down.
6.
Manually, adjust the vertical position of the camera until the image on the
LCD screen is sharp and in focus. Then, without moving the camera,
retighten the camera collar firmly.
Important:
After tightening the camera collar, verify that the image on the
LCD screen is still in focus. If it is not, repeat Steps 5 and 6 until the image
remains in focus.
7.
Reinstall the camera cover.
This completes the procedure for focusing the video camera on your Cobra
system.
B-2
Focusing and Aligning the Camera
Figure B-1. Focusing the Camera