2.8 FREQuency:LO1:EXTernal [1
|
0]
This command allows the user to select between the internal LO1 local oscillator and an externally sup-
plied one. Issuing this command overrides the back panel toggle switch setting.
When
0
is specified, the internal LO1 local oscillator is used by the Combined RF Upconverter/Downcon-
verter.
When
1
is specified, the external LO1 input is used by the Combined RF Upconverter/Downconverter.
Syntax
1
:
FREQ
:
LO1
:
EXT
Space
0
?
Example
:FREQ:LO1:EXT 0
This command selects the internal LO1 oscillator
.
Default Condition
On power up, or when a
*RST
command is issued, the QM1004-0.5-18 returns to the boot state’s setting.
The Combined RF Upconverter/Downconverter default setting is 0.
Query
:FREQ:LO1:EXT?
This query returns a 0 or 1 to indicate the selected Combined RF Upconverter/-
Downconverter reference mode
∙
0 is returned if the internal LO1 oscillator is selected
∙
1 is returned if the external LO1 is selected
Error Message
If the parameter is not in the recognized format, error
-102, "Syntax error"
occurs.
If the parameter is not 1 or 0, error
-222, "Data out of range"
occurs.
Quonset Microwave
Revision 1.2.1
41
Содержание QM1004-0.5-18
Страница 7: ...QM1004 0 5 18 Combined RF Upconverter Downconverter User Manual Quonset Microwave Revision 1 2 1 vi...
Страница 15: ...QM1004 0 5 18 Combined RF Upconverter Downconverter User Manual Quonset Microwave Revision 1 2 1 4...
Страница 40: ...QM1004 0 5 18 Combined RF Upconverter Downconverter User Manual 6 14 WAI 89 Quonset Microwave Revision 1 2 1 29...
Страница 105: ...Quonset Microwave www quonsetmicrowave com 1004 011 21 1004 011 21 Revision 1 2 1 November 2022...