
MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
306
Unable to set System Audio
SHUTDOWN
RESTART
* May not be able to adjust
audio
volume level
* Call service after multiple
occurrences
307
Low Arrest volume
< 10 ml
REFILL
CONFIRM
* Refill Arrest
308
Low Additive volume
< 10 ml
REFILL
CONFIRM
* Refill Additive
309
Crystalloid source volume low
< 150 ml
REPLACE
CONFIRM
* Stop flow to replace
Crystalloid
bag and enter new volume
310
Crystalloid Source Volume low
< 50 ml
REPLACE
CONFIRM
* Stop flow to replace
Crystalloid
bag and enter new volume
311
Flow set to zero
Crystalloid Source Empty
REPLACE
CONFIRM
* Stop flow to replace
Crystalloid
bag and enter new volume
312
Flow range is limited in Low Vol mode
Shift to NORMAL flow for full range
NORMAL
LOWVOL
* Select LOWVOL to remain in
Low Volume mode and stay
below 200 ml/min
* Select NORMAL to exit
LOWVOL mode and increase
the flow range
313
EEPROM Version Change Detected.
EEPROM Reset to Defaults
CONFIRM
* Manually Restart System
* Call service after multiple
occurrences
314
Different Console-Controller pair
detected
Press SWAP if intended, else
Press SHUTDOWN and reconnect
original
SWAP
SHUTDOWN
* Select SWAP to continue
OR
* Select SHUTDOWN to power
off system and re-connect
original pair
Alar
m
Cod
e
Message Display
Corrective Action
39
Содержание MPS 3
Страница 1: ......
Страница 4: ...This page intentionally blank...
Страница 6: ...This page intentionally blank...
Страница 10: ...MPS 3 and MPS 3 ND Troubleshooting Manual System Overview Quest Medical Inc 904678 Rev B 04 2023 4...
Страница 14: ......