MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
116
Arrest Cartridge detected in the Additive
Pump. Install the Arrest cartridge in the
Arrest Pump
* Check cartridges to ensure
Arrest Cartridge is inserted
into
Arrest Pump and Additive
Cartridge is inserted in
Additive
Pump
* If alarms persists when
Additive
cartridge is inserted into
Additive pump, remove
cartridge and try again
* If problem still exists after
replacing Additive cartridge,
restart console and try again
* Contact service after multiple
occurrences
117
Arrest pump error
Arrest pump is Off
CONFIRM
RETEST
* Select COFIRM to turn Arrest
pump off and continue
procedure
OR
* Select RETEST to turn Arrest
pump on and continue
procedure
118
Arrest pump failed
Proceed without Arrest pump
CONFIRM
* Restart System and resume
case
* Contact service after multiple
occurrences
119
Additive pump error
Additive pump is Off
CONFIRM
RETEST
* Select COFIRM to turn
Additive
pump off and continue
procedure
OR
* Select RETEST to turn
Additive
pump on and continue
procedure
Alar
m
Cod
e
Message Display
Corrective Action