MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
120
Additive pump failed
Proceed without Additive pump
CONFIRM
* Restart System and resume
case
* Contact service after multiple
occurrences
121
Arrest pump is Disabled
CONFIRM
* Prime arrest pump
122
Additive pump is Disabled
CONFIRM
* Prime additive pump
123
Circulation valve sensor mismatch
Manually monitor delivery temperature
H2O Circ is Off. Heaters are disabled.
CONFIRM
* Continue procedure without
ability to control temperature
using the console
* Try connecting medical air
* Restart System
* Contact service after multiple
occurrences
124
H2O temperature sensor error
H2O temperature will not be displayed
Heaters Disabled
CONFIRM
* Continue without temperature
control
OR
* Restart and resume case
* Call service after multiple
occurrences
125
Deliver temperature sensor error
Deliver temperature will not be displayed
Call Service after this Case
CONFIRM
* Manually monitor delivery
temperature
* Restart and resume case
* Call service after multiple
occurrences
126
Temperature exceeded allowed limit
Ensure that cold water is available
H2O Circ turned OFF
CONFIRM
* Check cold water source
temperature
* Ensure cold water source has
adequate water level and flow
rate
* Check blood inlet temperature
Alar
m
Cod
e
Message Display