MPS 3 and MPS 3 ND Troubleshooting Manual – Troubleshooting
5301016
MPS® 3 Blood Bypass Tubing
on
The MPS® 3 Blood Bypass Tubing contains the following component:
1.
MPS
®
3 Blood Bypass Tubing
on For Use
The MPS 3 Blood Bypass Tubing is only for use as a back-up with the
Quest MPS 3 Console to conZnue fluid delivery in the event of
complete ba1ery depleZon or if the MPS 3 Console becomes
unusable. The Blood Bypass Tubing, when used in conjuncZon with
the MPS Console and Delivery Set, is intended for use by perfusionists
and surgeons trained in the delivery of cardioplegia soluZons to the
myocardium during open heart surgery. This is a sterile single use
product.
Contron
This device is designed for use with only the Quest MPS 3 Console.
Use with other systems is contraindicated. This device is not designed,
sold or intended for use except as indicated.
Warnings and Cau+ons
1.
Read and understand the informaZon in these instrucZons
prior to operaZng the MPS 3 Console. See MPS 3
OperaZons Manual for a complete lisZng of Warnings and
CauZons and Storage CondiZons. The a1ending clinician is
solely responsible for the setup and use of the MPS 3
Console, Delivery Set, and Blood Bypass Tubing.
2.
Observe all addiZonal warnings and cauZons contained in this
InstrucZon for Use.
3.
Examine sterile package carefully before opening to confirm
package integrity and verify that the expiraZon date has not
passed. The device is supplied in a sterile, single use package and is
non-pyrogenic. DO NOT USE Blood Bypass Tubing with damaged
packaging or if the expiraZon date has passed. DO NOT resterilize
or reprocess.
4.
Observe asepZc technique with all tubing connecZons. Do not
overZghten rigid connecZons.
5.
The main door must be closed to operate the MPS 3 Console with
the Blood Bypass Tubing.
6.
Dispose of this device according to hospital procedure for
contaminated material.
Installing the Blood Bypass Tubing
1.
Using sterile technique, remove the Blood Bypass Tubing from the
sterile pouch.
2.
Ensure the clamp on the Blood Bypass Tubing is closed prior to
installaZon.
3.
Stop flow and ensure both delivery valves are closed.
4.
Clamp both Blood and Crystalloid inlet lines leading to main
pumping casse1e above the large bore luer connectors.
5.
Vent heat exchanger by manually depressing the vent valve to
reduce the system pressure to zero.
6.
Disconnect and cap blood source line from disposable casse1e
using female cap on Blood Bypass Tubing.
7.
A1ach Blood Bypass Tubing large bore luer connector to blood
source line.
8.
Connect Blood Bypass Tubing standard male luer connector to the
auxiliary luer port of 16 convoluZon heat exchanger or disconnect
addiZve check valve
and connect Blood Bypass Tubing onto lej port of 10 convoluZon
heat exchanger.
9.
Unclamp Blood Bypass Tubing and blood source line.
10.
Manually depress vent valve to prime Blood Bypass Tubing and
heat exchanger of air.
11.
Once primed, clamp Blood Bypass Tubing and release vent valve
12.
As protocols requires, unclamp Blood Bypass Tubing and then
iniZate flow by manually depressing delivery valve.
13.
If Controller is unresponsive, monitor pressure from Arterial
Monitor.
14.
Monitor heat exchanger and delivery line for air bubbles and vent
heat exchanger as needed.
Arrest Agent and Addi+ve Delivery with 16 on Heat
Exchanger
1.
Ensure delivery valves are closed and Blood Bypass Tubing is
clamped.
2.
Manually vent heat exchanger to reduce the system pressure to
zero.
3.
Connect syringes to arrest agent and addiZve stopcocks.
4.
Place arrest agent and addiZve stopcocks in the refill posiZon and
evacuate cartridge content.
5.
Turn and break arrest agent and addiZve stopcocks towards
cartridges.
6.
As protocol requires, inject arrest agent and/or addiZve cartridge
content into drug delivery lines.
7.
Unclamp Blood Bypass Tubing and iniZate flow with drug bolus by
pressing down on delivery valve.
Arrest Agent and Addi+ve Delivery with 10 on Heat
Exchanger
1.
Ensure delivery valves are closed and Blood Bypass Tubing is
clamped.
2.
Manually vent heat exchanger to reduce the system pressure to
zero.
3.
Connect syringes to arrest agent and addiZve stopcocks.
4.
Place arrest agent and addiZve stopcocks in the refill posiZon and
evacuate cartridge content.
5.
Turn and break arrest agent stopcock towards cartridge.
6.
Disconnect and connect arrest agent or addiZve syringe to yellow
arrest stopcock as protocol requires.
7.
Inject arrest agent or addiZve cartridge content into drug delivery
line as protocol requires.
8.
Unclamp Blood Bypass Tubing and iniZate flow and drug bolus by
pressing down on delivery valve.
Opera+ng ons
1.
The MPS 3 Delivery Set should not be used if there is
any visible damage.
2.
User should first clamp both inlet lines before disconnecZng from
pump casse1e.
3.
User is responsible for pressure monitoring as indicated by
hospital protocol.
4.
User is responsible for delivering proper arrest agent and addiZve
concentraZons by hospital protocol.
5.
User is responsible for purging air from the system as indicated by
hospital protocol.