2. Instructions for Proper Use 

i. 

Remove the existing mattress from the bed frame 

ii. 

Replace the standard mattress with mattress replacement system (orient mattress so that the air tube is at the 
foot of the bed) 

iii. 

Secure straps beneath the mattress to the bed frame 

iv. 

Hang the control unit on the foot board of the bed frame 

v. 

Attach the air tube connectors to the socket on the left panel of the control unit (connectors and sockets are 
colour coded) 

vi. 

Verify that air hoses are not kinked under the mattress 

vii. 

Attach cover to mattress 

viii. 

Plug in the control unit and turn on the master power switch on the right side panel (the STANDBY LED will 
illuminate) 

ix. 

Push the STANDBY/OPERATE button on the front panel (OPERATE LED will now be illuminated and the 
control unit will be in operation) 

x. 

Push the AUTO FIRM button for fast inflation. Allow 4-7 min for full inflation. 

xi. 

After the mattress is fully inflated, the caregiver can transfer the patient to the mattress (note: the mattress can 
be inflated while a patient is laying on it)  

xii. 

Push AUTO FIRM again to release the fast inflation mode 

xiii. 

Alternation time can be adjusted by the CYCLE TIME button. 

xiv. 

When the control unit is powered-on, it will default to previous setting. 

 

Static Function

: Push the STATIC button and adjust the comfort control by pressing the SOFT/FIRM button to 

achieve maximum patient comfort. On this mode the system provides low air loss therapy. Perform a hand 
check by placing 

hand under the patient’s buttocks between the cells and foam. The patient should have at 

least 4 cm of clearance between the buttocks and the bottom of the mattress. If the STATIC function is 
selected the time display will remain blank. 

 

 

Dynamic Function: 

Push the alternate button to enable the 3:1 alternating function. 

 

 

 

Lock-out Function: 

When lock-out function activates, buttons on control panel will be non-functional. 

 

CPR Deflation

: For quick mattress deflation disconnect the hose connector from the controller and release the 

 

CPR quick deflation valve. 

 

Содержание DynaLAL

Страница 1: ...User Manual DynaLAL Dynamic Low Air Loss DynaLAL Alternating Pressure Relief User Manual Distributed by Quart Healthcare Inc www quarthealthcare com ...

Страница 2: ...oduct Description The DynaLAL system is a unique and innovative specialized mattress replacement unit The system utilizes low air loss technology with a high flow rate that provides pressure management for the treatment of pressure ulcers The advanced 3 1 alternating function also provides active prevention for pressure relief especially for those in acute care and long term care settings the cell...

Страница 3: ...ction stopping the alternating function and providing only low air loss therapy The auto firm function provides uniform firmness for nursing procedures Power failures produce an audio alarm for added safety The alarm can be disabled by pushing the ALARM RESET Button on the front panel Double insulation provides near silent operation Foot board mounting rack provides convenient placement on bed ...

Страница 4: ...5 Device Identifier Health Canada FM CRP0005 Size cm 122 W x203 L x24 H 48 x 80 x 9 5 Weight Kg 21 0 Kg Cells Material PU Cover Type Zipper cover with removable foam base Cover Material Nylon knitted fabric w PU coating finish Base Material Woven Polyester fabric w PVC backing Part No FC QHI0079 Device Identifier Health Canada FC CRP0079 Size cm 42 L x17 W x27 H Weight Kg 6 1 Phase Time 5 95 Rated...

Страница 5: ...s is fully inflated the caregiver can transfer the patient to the mattress note the mattress can be inflated while a patient is laying on it xii Push AUTO FIRM again to release the fast inflation mode xiii Alternation time can be adjusted by the CYCLE TIME button xiv When the control unit is powered on it will default to previous setting Static Function Push the STATIC button and adjust the comfor...

Страница 6: ...tic cleaning Pre wash cold Main wash 70C 10 min Extraction 2 min Cold rinses TUMBLE DRYING OR TUNNEL DRYING IS NOT RECOMMENDED Mattress cells can be wiped with a solution of sodium hypochlorite 1000ppm or any other non phenolic germicidal solution Note The control unit should also be cleaned weekly using a damp soft cloth and mild detergent Switch off the electrical supply to the pump and disconne...

Страница 7: ...the mattress and disconnect the air feed tubes All of the air will be expelled Starting at the head of the mattress roll towards the foot of the bed Use the base mounted straps to secure It is recommended that the following guidelines are used whenever the system is being stored or transported to another location Temperature limitations 5C to 60C Relative humidity 30 to 75 2 Maintenance and Troubl...

Страница 8: ... in materials and workmanship Our obligation under this warranty is limited to the repair of equipment returned to the place of purchase within 24 months of delivery date We agree to service adjust any equipment returned and to replace or repair any part that is proven to be a warranty defect at no charge This warranty excludes equipment damage through shipping tampering improper maintenance carel...

Отзывы: