Cleaning procedure for air filter 
 

1.  Air Filter is located at the back of the control unit, inside the carry handle. No tools are required to remove 

it. Air Filter must be removed for inspection at least once a month and washed under running tap water if 
necessary. Squeeze to remove excess water before reinstalling it. 
 

2.  Use a soft brush to remove dust and debris from the air-intake muffler prior to reinstalling the filter. 
 
3.  New filters are also available. 
 

 

(Do not use phenol based cleaning solutions) 
(Switch off the electrical supply to the pump and disconnect the power cord from the main supply before cleaning 
and inspection) 

 
4. Storage and Care 

 

Control Unit: 

 

Check the power cord and plug for abrasions and excessive wear

 

 

Plug in the unit and verify air flow from the hose connection ports

 

 

Place in plastic bag for storage

 

 

Mattress Replacement System

 

 

Check the air manifold for kinks or breaks.  Replace if necessary. 

 

Twist open the CPR plug at the head of the mattress and disconnect the air feed tubes.  All of the air 
will be expelled.  Starting at the head of the mattress roll towards the foot of the bed.  Use the base 
mounted straps to secure. 

 
It is recommended that the following guidelines are used whenever the system is being stored or transported 
to another location: 
 

Temperature limitations  

5C to 60C 

 

Relative humidity 

    

30% to 75% 

 
 

2.  Maintenance and Troubleshooting

 

No daily maintenance is required.  This equipment should only be serviced by authorized technical personnel.  
In case of minor problems refer to the following: 
 

Symptom 

Inspection Procedure 

Possible Solution 

The air is flowing out from the control unit but 
the mattress is not inflating 

1.  Is the power source correct? Improper voltage may 

cause the pump to function abnormally and damage 
the control unit. 

2.  Do the air tubes allow smooth airflow? Is an air tube 

kinked? 

3.  Is there any air leakage from the air cells? 
4.  Is there any air leakage from air tube between 

mattress and control unit?  

5.  Has the air tube connector been connected properly? 

1.  Use power regulator. 

 
 
 

2.  Adjust the air tubes to enable smooth 

air flow. 

3.  Replace with new air cells 
 
4.  Replace with new air tubes 
 
 

Содержание DynaLAL

Страница 1: ...User Manual DynaLAL Dynamic Low Air Loss DynaLAL Alternating Pressure Relief User Manual Distributed by Quart Healthcare Inc www quarthealthcare com ...

Страница 2: ...oduct Description The DynaLAL system is a unique and innovative specialized mattress replacement unit The system utilizes low air loss technology with a high flow rate that provides pressure management for the treatment of pressure ulcers The advanced 3 1 alternating function also provides active prevention for pressure relief especially for those in acute care and long term care settings the cell...

Страница 3: ...ction stopping the alternating function and providing only low air loss therapy The auto firm function provides uniform firmness for nursing procedures Power failures produce an audio alarm for added safety The alarm can be disabled by pushing the ALARM RESET Button on the front panel Double insulation provides near silent operation Foot board mounting rack provides convenient placement on bed ...

Страница 4: ...5 Device Identifier Health Canada FM CRP0005 Size cm 122 W x203 L x24 H 48 x 80 x 9 5 Weight Kg 21 0 Kg Cells Material PU Cover Type Zipper cover with removable foam base Cover Material Nylon knitted fabric w PU coating finish Base Material Woven Polyester fabric w PVC backing Part No FC QHI0079 Device Identifier Health Canada FC CRP0079 Size cm 42 L x17 W x27 H Weight Kg 6 1 Phase Time 5 95 Rated...

Страница 5: ...s is fully inflated the caregiver can transfer the patient to the mattress note the mattress can be inflated while a patient is laying on it xii Push AUTO FIRM again to release the fast inflation mode xiii Alternation time can be adjusted by the CYCLE TIME button xiv When the control unit is powered on it will default to previous setting Static Function Push the STATIC button and adjust the comfor...

Страница 6: ...tic cleaning Pre wash cold Main wash 70C 10 min Extraction 2 min Cold rinses TUMBLE DRYING OR TUNNEL DRYING IS NOT RECOMMENDED Mattress cells can be wiped with a solution of sodium hypochlorite 1000ppm or any other non phenolic germicidal solution Note The control unit should also be cleaned weekly using a damp soft cloth and mild detergent Switch off the electrical supply to the pump and disconne...

Страница 7: ...the mattress and disconnect the air feed tubes All of the air will be expelled Starting at the head of the mattress roll towards the foot of the bed Use the base mounted straps to secure It is recommended that the following guidelines are used whenever the system is being stored or transported to another location Temperature limitations 5C to 60C Relative humidity 30 to 75 2 Maintenance and Troubl...

Страница 8: ... in materials and workmanship Our obligation under this warranty is limited to the repair of equipment returned to the place of purchase within 24 months of delivery date We agree to service adjust any equipment returned and to replace or repair any part that is proven to be a warranty defect at no charge This warranty excludes equipment damage through shipping tampering improper maintenance carel...

Отзывы: