User Manual 

– DynaLAL 

 

 
 
 
 

 

 

Warning 

 

 

Connect the Master Control unit to an appropriate power source

 

 

Do not use the system in the presence of any flammable gases (such as anesthetic agents)

 

 

Keep the pump and mattress away from sources of liquid and open flames

 

 

Keep mattress away from sharp objects

 

 

The device is not AP/APG protected

 

 

Keep mattress system away from heating devices

 

 

Caution 

 

  Consult clinical instructor before use of mattress  

  Support surfaces should always be used in conjunction with a care plan that includes the turning/repositioning 

of the patient over a 24-hour period 

  The control unit should only be repaired by an authorized distributor 

  Do not drop the control unit or store it in direct sunlight or extreme cold conditions 

 

1.  Product Description

 

 

The DynaLAL system is a unique and innovative specialized mattress replacement unit. The system utilizes 
low air loss technology with a high flow rate that provides pressure management for the treatment of pressure 
ulcers. The advanced 3:1 alternating function also provides active prevention for pressure relief, especially for 
those in acute care and long term care settings (the cells inflate and deflate in a 3:1 cycle, meaning 2/3rds of 
the body is always supported at any one time). The DynaLAL system offers “deep-cell therapy”; whereby, the 
cells completely collapse providing “0” pressure at the point of deflation. The soft-firm adjustment allows the 
patient to adjust the firmness or softness of the surface for optimal comfort through 8 digital scales. The 
surface also has 2 inches of enclosed convoluted foam to provide extra protection and comfort for the patient 
in the event of a power failure and the mattress deflates. 

 

Содержание DynaLAL

Страница 1: ...User Manual DynaLAL Dynamic Low Air Loss DynaLAL Alternating Pressure Relief User Manual Distributed by Quart Healthcare Inc www quarthealthcare com ...

Страница 2: ...oduct Description The DynaLAL system is a unique and innovative specialized mattress replacement unit The system utilizes low air loss technology with a high flow rate that provides pressure management for the treatment of pressure ulcers The advanced 3 1 alternating function also provides active prevention for pressure relief especially for those in acute care and long term care settings the cell...

Страница 3: ...ction stopping the alternating function and providing only low air loss therapy The auto firm function provides uniform firmness for nursing procedures Power failures produce an audio alarm for added safety The alarm can be disabled by pushing the ALARM RESET Button on the front panel Double insulation provides near silent operation Foot board mounting rack provides convenient placement on bed ...

Страница 4: ...5 Device Identifier Health Canada FM CRP0005 Size cm 122 W x203 L x24 H 48 x 80 x 9 5 Weight Kg 21 0 Kg Cells Material PU Cover Type Zipper cover with removable foam base Cover Material Nylon knitted fabric w PU coating finish Base Material Woven Polyester fabric w PVC backing Part No FC QHI0079 Device Identifier Health Canada FC CRP0079 Size cm 42 L x17 W x27 H Weight Kg 6 1 Phase Time 5 95 Rated...

Страница 5: ...s is fully inflated the caregiver can transfer the patient to the mattress note the mattress can be inflated while a patient is laying on it xii Push AUTO FIRM again to release the fast inflation mode xiii Alternation time can be adjusted by the CYCLE TIME button xiv When the control unit is powered on it will default to previous setting Static Function Push the STATIC button and adjust the comfor...

Страница 6: ...tic cleaning Pre wash cold Main wash 70C 10 min Extraction 2 min Cold rinses TUMBLE DRYING OR TUNNEL DRYING IS NOT RECOMMENDED Mattress cells can be wiped with a solution of sodium hypochlorite 1000ppm or any other non phenolic germicidal solution Note The control unit should also be cleaned weekly using a damp soft cloth and mild detergent Switch off the electrical supply to the pump and disconne...

Страница 7: ...the mattress and disconnect the air feed tubes All of the air will be expelled Starting at the head of the mattress roll towards the foot of the bed Use the base mounted straps to secure It is recommended that the following guidelines are used whenever the system is being stored or transported to another location Temperature limitations 5C to 60C Relative humidity 30 to 75 2 Maintenance and Troubl...

Страница 8: ... in materials and workmanship Our obligation under this warranty is limited to the repair of equipment returned to the place of purchase within 24 months of delivery date We agree to service adjust any equipment returned and to replace or repair any part that is proven to be a warranty defect at no charge This warranty excludes equipment damage through shipping tampering improper maintenance carel...

Отзывы: