Chapter 1 Specifications
4
Models QT-740 & QT-750 - Service Manual
ELECTRICAL SPECIFICATIONS
Input Power:
115 VAC (±10%), 15 Amps, 50/60 Hz
230 VAC (±10%), 10 Amps, 50/60 Hz
Mode of Operation:
Model QT-740:
Continuous
Model QT-750:
Intermittent Operation (80 cycles/hour max.)
Momentary Current:
Model QT-740:
2
.
0
Amps
Model QT-750: 5
.0
Amps
Long Term Current:
Model QT-740:
0.05 Amps
Model QT-750:
0.20 Amps
Locking System Power:
24 VDC/4.35 Amp
SYSTEM OPTIONS
•
Abdominal compression band
•
Lateral cassette holder (for cross-table work)
SYSTEM OPERATING ENVIRONMENT
Ambient Temperature:
+10°C to +40°C
Relative Humidity:
20 to 80%, non-condensing
Altitude:
-30.5 to +2440 meters relative to sea level
NON-OPERATING ENVIRONMENT
Ambient Temperature:
-18°C to +70°C
Relative Humidity:
20 to 95%, non-condensing
Altitude:
-30.5 to +3048 meters relative to sea level
Содержание QT-740
Страница 4: ...Revision History ii Models QT 740 QT 750 Service Manual THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 7: ...Chapter 1 1 SPECIFICATIONS...
Страница 8: ...2...
Страница 13: ...Chapter 7 2 ASSEMBLY INSTALLATION...
Страница 14: ...8...
Страница 29: ...Chapter 23 3 THEORY OF OPERATION...
Страница 30: ...24...
Страница 38: ...Chapter 3 Theory of Operation 32 Models QT 740 QT 750 Service Manual THIS PAGE INTENTIONALLY LEFT BLANK...
Страница 39: ...Chapter 33 4 SERVICE INSTRUCTIONS...
Страница 40: ...34...
Страница 57: ...Chapter 4 Service Instructions 51 Models QT 740 QT 750 Service Manual...
Страница 58: ...Chapter 51 5 DIAGRAMS...
Страница 59: ...52...
Страница 62: ...Chapter 5 Diagrams Models QT 740 QT 750 Service Manual 55 Figure 28 4 Way Table Control Board A1 AY40 017T...