5
Safety instructions
HBM
QuantumX
1
Safety instructions
Appropriate use
The module and the connected transducers may be used for measurement and directly
related control tasks only. Any other use is
not
appropriate.
To ensure safe operation, the module may only be used as specified in the operating
manual. It is also essential to follow the respective legal and safety regulations for the
application concerned during use. The same applies to the use of accessories.
Each time, before starting up the modules, you must first run a project planning and risk
analysis that takes into account all the safety aspects of automation technology. This
particularly concerns personal and machine protection.
Additional safety precautions must be taken in plants where malfunctions could cause major
damage, loss of data or even personal injury. In the event of a fault, these precautions
establish safe operating conditions.
This can be done, for example, by mechanical interlocking, error signaling, limit value
switches, etc.
Safety rules
A module must not be connected directly to a power supply network. The maximum
permissible supply voltage is 10 to 30 V (DC).
The supply connection, as well as the signal and sense leads, must be installed in such a
way that electromagnetic interference does not adversely affect device functionality (HBM
recommendation: ”Greenline shielding design”, downloadable from the Internet at
http://www.hbm.com/Greenline).
Automation equipment and devices must be covered over in such a way that adequate
protection or locking against unintentional actuation is provided (such as access checks,
password protection, etc.).
When devices are working in a network, these networks must be designed in such a way that
malfunctions in individual nodes can be detected and the nodes shut down.
Safety precautions must be taken both in terms of hardware and software, so that a line
break or other interruptions to signal transmission, such as via the bus interfaces, do not
cause undefined states or loss of data in the automation device.