80
| Troubleshooting
|
Ultra
Version M 22-April-2012
Interlock
Value
ICE Front Panel
Message
Possible Cause
Solution/Suggestion
Description
IF1 00001000
Power Supply
housing opened
Removal of one of the
ICE450 side covers
occurred.
•
Replace the cover.
If that does not resolve the issue, there may
be a faulty switch or an open circuit within
the ICE450. Call for technical assistance.
The ICE450 has a normally-open switch
installed on each side cover. When the side
covers are installed properly, the switches
are depressed.
IF1 00000100
controller bus error
Problem in the
controller system
relating to its internal
communications.
Contact Quantel Customer Service.
The ICE450 system controller has two
buses used to communicate with system
circuitry. A problem with either bus causes
an interlock fault condition.
IF1 00000001
flashlamp disabled
time out delay
expired
Flashlamp timeout has
expired.
Disable timeouts
or
change the flashlamp timeout to 00:00.
The flashlamp time-out interlock is a user-
defined parameter. Refer to the
for details.
IF2 10000000
heater over temp
Heater coolant is too
hot. The protective
thermal switch opened
when the coolant
neared 70°C.
•
Turn off power to the ICE450 and allow it
to cool down.
If this does not resolve the issue, the heater
thermostat circuit may need repair. Call
Quantel Customer Service.
A heater element (in some models)
warms the coolant to regulate coolant
temperatures in low-power applications.
A normally-closed thermal switch on the
heater block prevents cooling function
damage.
NOTE: To clear an interlock message, press the “Cursor Up”
or “Cursor Down” button on the Remote Box.