69
|
Ultra
Version M 22-April-2012
|
Ultra
7. Connect the empty Fill/Drain bottle. The tube from the
bottom of the bottle connects to the Fill/Drain port. The
tube from the top of the bottle connects to the Vent port.
8. Hold the bottle lower than the ICE450 reservoir. Loosen
the cap on the Fill/Drain bottle to allow the coolant to flow
freely.
Note:
Do not raise the bottle above the reservoir or the coolant
will flow back into the reservoir.
9. Once the reservoir is empty, use the Remote Box System
menu and change the Pump to Drain mode (see
Note:
The pump will run for 30 seconds and turn OFF
automatically.
10. After the pump stops automatically, tilt the ICE450 forward
toward the front panel to drain any remaining coolant into
the bottle.
11. Turn the Key Switch to OFF.
12. Disconnect the Fill/Drain bottle and the Coolant Loop from
the ICE450.
13. To prepare the system for shipment, wipe all residual
coolant droplets from all coolant ports prior to placing
the ICE450 in a plastic bag. This helps prevent moisture
condensing on sensitive parts during shipment.
14. Gravity-drain the Laser Head by tilting it so that coolant
flows out from the lower drain port and into your container.
15. Gently blow dry nitrogen through the blue coolant
connector to empty any remaining coolant.
16. Place the blue plastic covers that originally shipped with
the system over the ports on the Laser Head to prepare it
for shipping or storage.
Note:
The trace amount of coolant remaining in the system
after following this procedure is not a concern.
Fill/Drain port
faces down
Keep bottle
lower than
reservoir
Tilt the ICE450 so the front
panel is lower than the
rear panel to drain any
remaining coolant into the
Fill/Drain bottle.
Disconnect the tubing from
the Laser Head. Tilt Laser
Head to drain. Gently blow
dry nitrogen through the blue
coolant connector to empty any