38
| Software
|
Ultra
Version M 22-April-2012
Feature
Operation
Command
Response
format
Description
Q-Switch
Commands
Fire Q-Switch, set
Example: CC
fire auto qs
CC
state string
Starts the Q-Switching. The response to this command is the
state string.
Fire Q-Switch Single Shot
Example: OP
fire auto qs
OP
state string
Initiates a single Q-Switch. The response to this command is
the state string.
Note:
This functions only when Q-Switch mode is set to INT.
Stop Q-Switch
Example: CS
fire auto
CS
state string
Disables the Q-Switch. The response to this command is the
state string.
Q-Switch Delay, query
Example: W
delay 123 μs
Q-Switch Delay, set
Example: W144
delay 144 μs
W
Wnnn
where nnn is
the Q-Switch
delay in
microseconds
delay nnn μs
where nnn is the
Q-Switch delay in
microseconds
Queries or sets the Q-Switch delay setting. This is the
amount of time, in microseconds, between the flashlamp
activation and the Q-Switch pulse when the Q-Switch is in
Internal sync mode. The max-delay command (“WMX”) sets
the maximum allowable delay, and the min-delay command
(“WMN”) sets the minimum allowable delay.