Rev 10-2009
SERVICE MANUAL
P - 5
1 - TECHNICAL SPECIFICATIONS
- Power supply:
External module with automatic voltage adaptation,
no selection is needed.
Input Voltage:
110 to 230 Vac ± 10%
single phase without earth
Frequency:
50 or 60 Hz
Mains consumption:
20 VA
AXIS-II consumption:
5.4 W (450mA on 12Vdc)
Class II Power Supply:
Reinforced Isolation.
- References:
SSL20-7612
Trade Mark:
ARTESYN
Technologies
or
ESS36US12-1E
Trade Mark:
XP POWER
or
TR30M120-01E03
Trade Mark:
CINCON
- Compliance
:
Power supply and enclosure designed per
IEC 60601-1
European directive:
93/42/CEE
Electromagnetic compatibility:
IEC 60601-1-2
1-1 ELECTRICAL REQUIREMENTS
1-2 ENVIRONMENTAL CONDITIONS
The temperature of the room where the device is operated must be within the
following range:
50°F < T° < 95°F
(10 °C < T° < 35 °C)
The relative humidity must not exceed
95%
without condensation.
The Device storage and transport temperature must be within the range:
- 4°F < T° < 158°F
(-20 °C < T° < 70 °C)
1-3 HOUSING MATERIAL
Upper frame:
Polystyrene PS CHOC complies with UL 94 VO standard
Lower frame:
Painted steel sheet
LCD front panel:
Polycarbonate