![Quantel Medical Axis II PR Скачать руководство пользователя страница 32](http://html1.mh-extra.com/html/quantel-medical/axis-ii-pr/axis-ii-pr_service-manual_3460982032.webp)
Rev 10-2009
SERVICE MANUAL
P - 32
APPENDIX - A (4/8)
To upgrade a Video Interface Board First Generation into Second Generation,
follow the procedure as below:
a)
Cut the two tracks located between :
-
pin 11 of connector J1
and
pin 3 of the IC2 circuit
-
pin 12 of connector J1
and
pin 4 of the IC2 circuit
b)
Reverse the signal of the contrast by connecting a wire between :
-
pin 11 of connector J1
and
pin 4 of the IC2 circuit
-
pin 12 of connector J1
and
pin 3 of the IC2 circuit
To use a second generation video Interface Board with a first generation of LCD screen,
you can use the same wiring and change the way they are connected, as indicated
below :
See details as below
Video interface board
Cut the tracks Here
The tracks are cut
Connect new wires to
reverse the contrast signal :
New wire
New wire
Reverse the wires to use the video board
with a first generation of LCD screen :
Original connection for second
generation of LCD screen :