![Quantel Medical Axis II PR Скачать руководство пользователя страница 14](http://html1.mh-extra.com/html/quantel-medical/axis-ii-pr/axis-ii-pr_service-manual_3460982014.webp)
Rev 10-2009
SERVICE MANUAL
P - 14
3-5 REPLACING AN LCD SCREEN
Remove the 2
screws
1
2
Unplug the
cable
connectors
3
Pull the board in
this direction
to remove it from
the notches
4
5
To replace the board, perform the
steps above in reverse order
Warning
: the faulty board must be returned to QUANTEL MEDICAL as a
whole without unmounting or opening any component. Failure to comply
will void the equipment warranty.
3 - BOARD DESCRIPTION
Note
:Ribbon
blue mark is up
Note
: Before disconnecting the ribbon cable, note the polarizing position
(blue side up) for future reconnection. On the V5 board, the ribbon cable
can be disconnected from both end while on earlier version boards, the
cable can be removed only on one end.
Unplug the
ribbon cable
connector
Note
:Ribbon
blue mark is up
Board notches
XEAX2LCDCOLOR