Follow the steps below for attaching and detaching accessories:
1. Disconnect probe from convertor. Use t he w rench set provided w ith the system.
2. Clean t hreaded stud. Use alcohol and a cotton sw ab to remove any debris on the
threading of the connecting stud. Allow the alcohol t o dry completely.
3. Clean threading in converter. Hold the convertor horizontally and use alcohol and a
cotton sw ab to remove any debris on t he threading. Do not allow liquid to drip into
Convertor. Allow the alcohol t o dry completely.
4. Reattach probe to convertor. Screw the probe back ont o the convertor and tighten w ith
the w rench set provided.
5. The tips on replaceable tip probes can be removed for cleaning and/or replacement.
When replacing horns or horn tips, Alw ays clean the threaded mating surfaces of the
convertor and horn. Use alcohol and a cotton sw ab to remove any debris on the
threading of the tip or probe. Note: If the replaceable tip loosens during sonication, be
sure to remove the tip f or cleaning and inspect the threading on t he tip and probe. Call