Sequence Progress Bar
– The measure of the total of all programs in the sequence including
On, Off and Delay After times.
The follow ing Option Screen can be accessed from either Menu Screen.
The Option menu enables the selection of
Temperature units
and
Footswitch
operational
mode
.
Option Screen
Temperature Units
– A temperature probe (Part# 4102 or 4103) is required for use of this
option.
The Temperature Probe option can be selected if you w ish to monitor t he temperat ure of
the sample being processed.
The temperature probe must be plugged into the back of the ultrasonic processor and the
probe tip must be in t he liquid sample. If the initial sample temperature is above the set
point, sonication w ill not turn on. During sonication, if the temperature reaches the
Shutdow n value, t he unit w ill pause sonication. The unit w ill remain paused until the
temperature drops below the pre-set temperat ure limit , at w hich time sonication can be
manually resumed.
Footswitch Operation
– A footsw itch (# FS-3) is required for use of this option. After
setting up a program, the footsw itch can be used to remotely activate the unit.
Exit
– The Exit button w ill save changes and return to the previous screen.