Page
15
of
30
7. OPERATING INSTRUCTIONS
7.1 CAUTION
Do not operate the power supply unless it is connected to the converter.
Never allow liquid to spill into the converter.
Do not allow a Microtip to vibrate in air.
Do not allow the vibrating Microtip to contact anything but the sample.
Never place a washer between the converter, probe or horn.
Never apply grease to the mating surfaces or threads of any component.
Should it become necessary to remove a probe, use the wrenches supplied. Never attempt to
remove the probe by twisting the converter housing or holding it in a vice, as this may
damage the electrical connections within the housing.
Overheating will damage the converter. If continuous operation for more than 15 minutes is
required, see addendum for converter cooling instructions.
The converter should remain near room temperature at all times. The sample liquid
temperature should not exceed
60
o
C (140
o
F)
. If either of these temperature limits are
reached, shut the system off and allow to cool.
Replaceable tip probes (1/2” – 1” diameter) are made for use with water based samples only.
If the liquid being processed is an organic solvent or any liquid with a lower surface tension
than water – always use a Solid tip probe. Solid tips can be used with any type of liquid.
Replaceable tip
Solid tip
Содержание Sonicator Q500
Страница 2: ...Page 2 of 30 ...