QTS-6000 Series
Analog Transmitter/Sensor
B71050-021-000-RB.doc
ELECTROCHEMICAL TRANSMITTER
JUL. 18, 2007
PRODUCT MANUAL
14
8.2.1 I
NDUSTRIAL
E
XPLOSION
-P
ROOF
E
NCLOSURES
A) Remove cap and screen from sensor assembly.
B) Remove old sensor element, including O-ring seal.
C) Install new sensor element, including O-ring seal. Take care to align the pins to the socket.
D) Replace cap and screen on sensor assembly.
8.2.2 ABS P
LASTIC
C
OMMERCIAL AND
NEMA 4X E
NCLOSURES
A) Remove four mounting screws located one at each corner of the front cover.
B) Securing the transmitter assembly from falling, remove the three transmitter mounting screws from the
front cover. Transmitter, sensor element, and grounding ring will all come free.
C) Remove old sensor element from transmitter pin sockets.
D) Install new sensor element into transmitter pin sockets. Take care to align the pins to the socket.
E) Remove O-ring from new sensor element.
F) Re-install sensor/transmitter against grounding ring on the inside front cover. Secure with three
mounting screws through the front cover.
G) Secure front cover with four cover screws, one on each corner.
Figure 10 - Sensor Element Installation
ABS Plastic Commercial or NEMA 4X Enclosures
Figure 9 - Sensor Element Installation