8.0 Taking readings
8.1 Preliminary checks before taking readings (best practice)
Prior to use, it is good practice to ensure
that:
Step 1
If using LSGAM ID codes and
information have been uploaded
from LSGAM to the analyzer.
Please see section 8.1.1 for more
information on this.
Step 2
The instrument has the correct
time and date set.
Step 3
The water trap filter is fitted and
is clean and dry.
Step 4
The battery has a good charge
(minimum 25% charge, even if
only a few readings are required).
Step 5
All gas channels have been
zeroed, without gas concentration
present.
Step 6
If necessary check the span
calibration with a known
concentration calibration gas.
Step 7
Take readings.
Start
ID codes
and
information
uploaded
Analyzer has
correct time and
date set
Water trap filter is
clean and dry
Battery has
good charge
Gases
zeroed
Check span
calibration
Take readings
Step 1
Step 2
Step 6
Step 5
Step 4
Step 3
Step 7