Guidance and manufacturer’s declaration-electromagnetic immunity
The B100 is intended for use in the electromagnetic environment specified below.
The customer or the user of the B100 should assure that it is used in such an environment
Immunity test
IEC 60601 test level
Compliance level
Electromagnetic environment - guidance
Electrostatic discharge(ESD)
IEC 61000-4-2
± 6 kV contact
± 8 kV air
± 6 kV contact
± 8 kV air
Floors should be wood, concrete or ceramic tile.
If floors are covered with synthetic material,
the relative humidity should be at least 30%
Electrical fast
transient/burst IEC
61000-4-4
± 2kV for power supply lines
+ 1kV for input/output lines
N/A
Mains power quality should be that of a typical commercial
or hospital environment.
Surge IEC 61000-4-5
± 1kV line(s) to line(s)
± 2kV line(s) to earth
N/A
Mains power quality should be that of a typical commercial
or hospital environment.
Voltage Dips, short
interruptions and voltage
variations on power supply
input lines IEC 61000-4-11
5% UT(95% dip in UT) for 0,5 cycle
40% UT(60% dip in UT) for 5 cycles
70% UT(30% dip in UT) for 25 cycles
5% UT(95% dip in UT) for 5 s
N/A
Mains power quality should be that of a typical commercial
or hospital environment. If the user of the B100 requires
continued operation during power mains interruptions, it is
recommended that the B100 be powered from an
uninterruptible power supply or a battery.
Power frequency (50, 60 Hz)
magnetic field IEC
61000-4-8
3 A/m
3 A/m
The B100 power frequency magnetic fields should be at
levels characteristic of a typical location in a
typical commercial or hospital environment.