Harmonic emissions
IEC 61000-3-2
Not applicable
The BP-801 Wireless Blood Pressure Monitor is
suitable for use in all establishments, including
domestic establishments and those directly
connected to the public low-voltage power supply
network that supplies buildings used for domestic
purposes.
Voltage fluctuations/
Flicker emissions
IEC 61000-3-3
Not applicable
Declaration – electromagnetic emissions and immunity for equipment and systems that are not life-
supporting and are specified for use only in a shielded location
The Wireless Blood Pressure Monitor declaration-
electromagnetic immunity
Declaration – electromagnetic immunity
The Wireless Blood Pressure Monitor system is
intended for use in the electromagnetic environment
specified below. The customer or the user of the
Wireless Blood Pressure Monitor system should
assure that it is used in such an environment.
The Wireless Blood Pressure Monitor system is
intended for use in the electromagnetic environment
specified below.
The customer or the user of the Wireless Blood
Pressure Monitor system should assure that it is used
in such an environment.
Immunity
test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Immunity
test
IEC
60601
test level
Compliance
level
Electromagnetic
environment -
guidance
Electrostatic
discharge
(ESD)
IEC 61000-4-
2
6 kV
contact
8 kV air
6 kV
contact
8 kV air
Floors should be
wood, concrete
or ceramic tile.
If floors are
covered with
synthetic
material, the
relative humidity
should be at least
30 %.
Conducted
RF IEC
61000-4-6
3 Vrms
150 kHz
to 80
MHz
N/A
Portable and
mobile RF
communications
equipment should
be used no closer
to any part of the
EQUIPMENT or
SYSTEM including
cables, than the
recommended
separation
distance
calculated from
the equation
applicable to the
frequency of the
transmitter.
Interference may
occur in the
vicinity of
equipment
marked with the
following symbol.
Electrical
fast
transient/
Burst IEC
61000-4-4
2 kV for
power
supply
lines 1 kV
for input
/output
lines
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Surge IEC
61000-4-5
1 kV
differenti
al mode
2 kV
common
mode
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
environment.
Radiated
RF IEC
61000-4-3
3 V/m
80 MHz
to 2.5
GHz
N/A
Voltage dips,
short
interruption
s and
voltage
variations on
power
supply input
lines IEC
61000-4-11
-5 % UT
(95 %
dip in
UT) for
0.5 cycle
, -40 %
UT (60 %
dip in
UT) for 5
cycles
-70 % UT
(30 %
dip in
N/A
Mains power
quality should be
that of a typical
commercial or
hospital
environment. If
the user of the
EQUIPMENT or
SYSTEM requires
continued
operation during
power mains
interruptions, it is