Improper continuous pressurization from cuff’s inflation on the same limb will cause loss of function 
of device.  

 

Do not apply the cuff over the users’ arm bearing a wound or medical treatment and so on, as this 
can cause further influence on the therapy. 

 

When cuff inflation, always check the device will not result in prolonged impairment of the 
circulation of the blood on users.  

 

Do not disassemble the QardioArm blood pressure monitor. 

 

Do not use an alcoholic-based or solvent agent to clean the device. Clean the device only with a soft, 
dry cloth. Clean the cuff carefully with a humid cloth and soap. Do not submerse any part of the 
QardioArm in water at any time. 

AAA alkaline cells usage, storage 

If AAA alkaline batteries fluid should get on your skin or clothing, immediately and thoroughly rinse with 
plenty of clean water. 

Use only four AAA alkaline batteries with the QardioArm blood pressure monitor. Do not use any other 
types of AAA batteries. 

When replacing batteries, insert the four AAA alkaline batteries with their polarities aligned as indicated 
on the QardioArm blood pressure monitor. 

Immediately replace the AAA alkaline batteries when they are depleted. 

Always replace all four AAA alkaline batteries at the same time: do not use new and old AAA alkaline 
batteries together. 

If the QardioArm blood pressure monitor will not be used for a long period of time, the batteries should 
be removed. 

Store the device and the components in a clean, dry and safe location 

Error messages and troubleshooting 

Problem 

Cause 

Remedy 

START button is gray, 
not green 

QardioArm is not connected to 
your smartphone or tablet. 

1) Close the QardioArm cuff and reopen it again.  
2) Ensure that Bluetooth is enabled on your phone or 
tablet and the QardioArm is nearby your phone or 
tablet.  
3) Replace the batteries of the QardioArm. Reset the 
pairing. 

Measurement could 
not be performed 

During the measurement, error 
signals were auto-detected by 
the device, or the pulse signals 
on the cuff are too weak.. 

Re-position the cuff and repeat the measurement 
keeping your arm still and without talking. Check the 
marks on the cuff and the instructional videos on the 
Qardio App to verify you are positioning the 
QardioArm correctly. If the problem occurs again, 
please contact customer service. 

No pressure in the 
cuff 

An adequate pressure cannot 
be generated in the cuff. A leak 
may have occurred. 

1) Check that the cuff is correctly positioned and fit to 
the arm.  
2) Replace the batteries if necessary.  
Repeat the measurement. If the problem occurs 
again, please contact customer service. 

Abnormal result 

The measuring signals are 
inaccurate and no result can 
therefore be displayed. This 
could be due to cuff not fully 

1) Read through instructions for performing reliable 
measurements. Re-position the cuff and repeat the 
measurement keeping your arm still and without 
talking.  If the problem occurs again, please contact 

Содержание QardioArm A100

Страница 1: ...QARDIO QardioArm A100 A Smarter Read on Blood Pressure User Guide Draft 0 9 drawings to be replaced not paginated and not formatted...

Страница 2: ...is device is not intended to be a diagnostic device Contact your physician if hypertensive values are indicated The cuff circumference is limited to 22cm 37cm 8 7in 14 6in Important safety information...

Страница 3: ...nder the cuff 2 Always ensure that the cuff fits correctly verifying the markings on the cuff You should fit the cuff closely but not too tight 3 Make sure that the cuff is positioned 2cm 0 8in above...

Страница 4: ...re the measurement Always measure on the same arm normally left or as instructed by your doctor Ideally remove close fitting garments from the upper arm or ensure that rolling the sleeve up does not c...

Страница 5: ...subject to wide fluctuations throughout the day depending on various potential factors including physical exertion and condition You should routinely take your measurements in quiet conditions when yo...

Страница 6: ...mpartment 3 Put the batteries compartment hatch back in place pushing until it clicks in place 4 You will see a green light shining through the batteries compartment hatch Resetting the pairing In ord...

Страница 7: ...pressure Do not take measurement immediately after bathing drinking alcohol smoking exercising or eating Do not use the QardioArm blood pressure monitor while operating a mechanical vehicle or in a mo...

Страница 8: ...ed for a long period of time the batteries should be removed Store the device and the components in a clean dry and safe location Error messages and troubleshooting Problem Cause Remedy START button i...

Страница 9: ...and correct if required 3 Replace the AAA alkaline cells Each measurement has significantly different results Under normal measuring circumstance the reading at home is different from that of the clin...

Страница 10: ...user must deliver it to the able collecting centers for electric and electronic garbage or give back to the retailer when purchasing a new device Disposing of the product separately prevents possible...

Страница 11: ...gnetic Compatibility FCC Rule Part 15 247 Cat DSS Bluetooth FCC Rule Part 15 247 Cat DTS BT4 0 EN ISO 10993 1 2009 Biological evaluation of medical devices Part 1 Evaluation and testing within a risk...

Страница 12: ...s must follow the specific operating instructions for satisfying RF exposure compliance This transmitter must not be co located or operating in conjunction with any other antenna or transmitter Indust...

Страница 13: ...uld be used no closer to any part of the device including cables than the recommended separation distance calculated from the equation applicable to the frequency of the transmitter The device is inte...

Страница 14: ...l Compliance level Electromagnetic environment guidance Electrostatic discharge ESD IEC 61000 4 2 6 kV contact 8 kV air 6 kV contact 8 kV air Floors should be wood concrete or ceramic tile If floors a...

Страница 15: ...istered in the U S and other countries The Bluetooth word mark and logos are registered trademarks owned by Bluetooth SIG Inc and any use of such marks by Qardio Inc is under license Other trademarks...

Отзывы: