3.2
Optional Accessories
Figure 2.
4
Installation
4.1
Battery Installation
The LSP Series sensor can be powered by four (4) AA Li/SOCl
2
batteries if a 24 VDC supply is not available.
These batteries are unique and available through Pyxis Lab (P/N: 50731). Typical battery life after replac-
ing a new battery set is about 6 months with a 60 minute measurement interval. The LSP Series battery
compartment is shown in Figure 3.
*
NOTE
*
Do
NOT
use rechargeable nickel cadmium (NiCad) or rechargeable lithium batter-
ies.
Figure 3. Battery Installation
LSP Series User Manual
| +1 (866) 203-8397
4